AUTHORS: Celicanin Z, Auboiroux V, Bieri O, Petrusca L, Santini F, Viallon M, Scheffler K, Salomir R
Magnetic Resonance in Medicine, 72(4): 1087-95, October 2014
ABSTRACT
PURPOSE:
Magnetic resonance-guided high-intensity focused ultrasound is considered to be a promising treatment for localized cancer in abdominal organs such as liver, pancreas, or kidney. Abdominal motion, anatomical arrangement, and required sustained sonication are the main challenges.
METHODS:
MR acquisition consisted of thermometry performed with segmented gradient-recalled echo echo-planar imaging, and a segment-based one-dimensional MR navigator parallel to the main axis of motion to track the organ motion. This tracking information was used in real-time for: (i) prospective motion correction of MR thermometry and (ii) HIFU focal point position lock-on target. Ex vivo experiments were performed on a sheep liver and a turkey pectoral muscle using a motion demonstrator, while in vivo experiments were conducted on two sheep liver.
RESULTS:
Prospective motion correction of MR thermometry yielded good signal-to-noise ratio (range, 25 to 35) and low geometric distortion due to the use of segmented EPI. HIFU focal point lock-on target yielded isotropic in-plane thermal build-up. The feasibility of in vivo intercostal liver treatment was demonstrated in sheep.
CONCLUSION:
The presented method demonstrated in moving phantoms and breathing sheep accurate motion-compensated MR thermometry and precise HIFU focal point lock-on target using only real-time pencil-beam navigator tracking information, making it applicable without any pretreatment data acquisition or organ motion modeling.
Download PDF
BibTex
Module: MRI