AUTHORS: Meyerspeer M, Magill AW, Kuehne A, Gruetter R, Moser E, Schmid AI

Magnetic Resonance in Medicine, 76(5): 1636–1641, November 2016


ABSTRACT

Purpose

Modification of a clinical MRI scanner to enable simultaneous or rapid interleaved acquisition of signals from two different nuclei.

Methods

A device was developed to modify the local oscillator signal fed to the receive channel(s) of an MRI console. This enables external modification of the frequency at which the receiver is sensitive and rapid switching between different frequencies. Use of the device was demonstrated with interleaved and simultaneous 31P and 1H spectroscopic acquisitions, and with interleaved 31P and 1H imaging.

Results

Signal amplitudes and signal‐to‐noise ratios were found to be unchanged for the modified system, compared with data acquired with the MRI system in the standard configuration.

Conclusion

Interleaved and simultaneous 1H and 31P signal acquisition was successfully demonstrated with a clinical MRI scanner, with only minor modification of the RF architecture. While demonstrated with 31P, the modification is applicable to any detectable nucleus without further modification, enabling a wide range of simultaneous and interleaved experiments to be performed within a clinical setting.