AUTHORS: Yu J, Agarwal H, Stuber M, Schar M

Journal of Magnetic Resonance Imaging, 33(6): 1330-40, June 2011


ABSTRACT

Purpose: 

To develop and evaluate a practical method for the quantification of signal-to-noise ratio (SNR) on coronary MR angiograms (MRA) acquired with parallel imaging.

Materials and methods: 

To quantify the spatially varying noise due to parallel imaging reconstruction, a new method has been implemented incorporating image data acquisition followed by a fast noise scan during which radiofrequency pulses, cardiac triggering and navigator gating are disabled. The performance of this method was evaluated in a phantom study where SNR measurements were compared with those of a reference standard (multiple repetitions). Subsequently, SNR of myocardium and posterior skeletal muscle was determined on in vivo human coronary MRA.

Results: 

In a phantom, the SNR measured using the proposed method deviated less than 10.1% from the reference method for small geometry factors (≤2). In vivo, the noise scan for a 10 min coronary MRA acquisition was acquired in 30 s. Higher signal and lower SNR, due to spatially varying noise, were found in myocardium compared with posterior skeletal muscle.

Conclusion: 

SNR quantification based on a fast noise scan is a validated and easy-to-use method when applied to three-dimensional coronary MRA obtained with parallel imaging as long as the geometry factor remains low.

Download PDF


BibTex


Module: