AUTHORS: Widmaier M., Kaiser A., Baup S., Xiao Y., Jiang Y., Huang Z., Wenz D., Xin L.

ISMRM & SMRT Annual Meeting & Exhibition, : , Singapore, May 2024


ABSTRACT

Motivation

Fast and accurate B1 mapping for 31P MRI is challenging due to the low sensitivity of 31P nucleus and its long T1 relaxation times.

Goal

This study aims to develop a fast B1 mapping approach for 31P at 7T.

Approach

A look-up table approach was adopted to enable B1 mapping at short TR. Fast spiral encoding together with weighted averaging was implemented with a GRE sequence to further accelerate the acquisition.

Results

B1 mapping was validated in phantom and in vivo in human calf muscle and brain within a scan time of 10min.

Impact

The presented fast and accurate B1 mapping correction method is particularly suited for moderate to short repetition times, showing promise for future applications in rapid X-nuclei imaging.